• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION EXPRESS SD RENAL/BILIARY; CATHETER, BILIARY, DIAGNOSTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION EXPRESS SD RENAL/BILIARY; CATHETER, BILIARY, DIAGNOSTIC Back to Search Results
Model Number 20188
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/14/2021
Event Type  malfunction  
Event Description
It was reported that shaft break occurred.The stenosed target lesion was located in the moderately tortuous and severely calcified iliac vein.A 6.0mmx14mmx90cm express sd stent was advanced but the shaft broke and lost push ability.During removal, the stent came off the catheter.The physician then used an opticross 18 intravascular ultrasound (ivus) to look on the stent but the ivus got stuck in the lesion and the screen went black.Subsequently, a 7.0mmx15mmx90cm express sd stent was used to tack up the smaller stent but it also broke during deployment and had to be removed.The procedure was completed using additional sd stent.No patient complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXPRESS SD RENAL/BILIARY
Type of Device
CATHETER, BILIARY, DIAGNOSTIC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key12263574
MDR Text Key264699787
Report Number2134265-2021-09631
Device Sequence Number1
Product Code NIN
UDI-Device Identifier08714729390442
UDI-Public08714729390442
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K152607
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 08/02/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/02/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/01/2024
Device Model Number20188
Device Catalogue Number20188
Device Lot Number0027030057
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/15/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/02/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-