• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER SURGICAL, INC. FAN SPRAY KIT; LAVAGE, JET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER SURGICAL, INC. FAN SPRAY KIT; LAVAGE, JET Back to Search Results
Model Number N/A
Device Problems Expulsion (2933); Power Problem (3010); Pressure Problem (3012)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/12/2021
Event Type  malfunction  
Manufacturer Narrative
This event has been recorded by zimmer biomet under (b)(4).Visual examination of the returned product identified that the device was filled with fluid.The handpiece and battery were soaked with fluid.The interior of the handgun was wet and rusted/corroded.The battery pack was wet on the inside and the plastic was warped.Device is used for treatment.A definitive root cause cannot be determined.The event cannot be confirmed.
 
Event Description
It was reported that the surgeon couldn't use the fan spray kit since saline water running from the device was unstable.Investigation found the batteries were corroded and peeling.No adverse event was reported as a result of this malfunction.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FAN SPRAY KIT
Type of Device
LAVAGE, JET
Manufacturer (Section D)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer (Section G)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer Contact
christina arnt
56 e. bell drive
warsaw, IN 46582
5745273773
MDR Report Key12266149
MDR Text Key264700613
Report Number0001526350-2021-00874
Device Sequence Number1
Product Code FQH
UDI-Device Identifier00889024375161
UDI-Public(01)00889024375161(17)230630(10)64755619
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 08/03/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/03/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/30/2023
Device Model NumberN/A
Device Catalogue Number00515047501
Device Lot Number64755619
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/13/2021
Was the Report Sent to FDA? No
Date Manufacturer Received08/02/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/03/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
-
-