Model Number L311
Device Problem Use of Device Problem (1670)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/30/2021
Event Type  Injury  
Event Description
It was reported that this patient with this pacemaker exhibited failure to capture on the right atrial (ra) lead.In addition, it was reported that while performing a threshold test through the programmer, telemetry issues were encountered.Troubleshooting measures were taken.A successful threshold test was achieved, through aai mode, but no capture was observed.Subsequently, additional intervention was taken to revise the ra lead.No additional adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACCOLADE MRI DR
Type of Device
IMPLANTABLE PULSE GENERATOR, PACEMAKER (NON-CRT)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key12278185
MDR Text Key265087521
Report Number2124215-2021-23366
Device Sequence Number1
Product Code LWP
UDI-Device Identifier00802526559228
UDI-Public00802526559228
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P150012/S000
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Non-Healthcare Professional
Remedial Action Patient Monitoring
Type of Report Initial
Report Date 08/04/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/04/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date04/29/2023
Device Model NumberL311
Device Catalogue NumberL311
Device Lot Number553350
Was Device Available for Evaluation? No
Date Manufacturer Received06/30/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/14/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age85 YR
-
-