• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TRUMPF MEDIZIN SYSTEME GMBH + CO. KG TRULIGHT 5520; SURGICAL LIGHT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TRUMPF MEDIZIN SYSTEME GMBH + CO. KG TRULIGHT 5520; SURGICAL LIGHT Back to Search Results
Model Number 4065520
Device Problems Degraded (1153); Flaked (1246); Material Integrity Problem (2978)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/03/2021
Event Type  malfunction  
Manufacturer Narrative
The investigation identified steps in the preparation for coating process which can lead to lack of paint adhesion resulting in paint chipping.Additionally, collisions of the light heads and use of unapproved cleaners can further contribute to paint chipping from the light system.The damage to the painted surfaces usually does not develop suddenly but develops over time.The instructions for use state that the devices must be checked for proper condition prior to each use.Damaged devices must not be used.If damage to the surface coating is recognized and removed, there is no danger to the patient.The light head and spring arm cover have been replaced and the device is functioning as designed.The surface coating process has been updated to include additional testing and verification.Based on this information, no further action is required.
 
Event Description
Trumpf medical received a report from the customer stating that paint particles are chipping from the spring arm and light head of the light system.No injuries were reported in relation to this allegation.This report was filed in our complaint handling system as complaint #(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRULIGHT 5520
Type of Device
SURGICAL LIGHT
Manufacturer (Section D)
TRUMPF MEDIZIN SYSTEME GMBH + CO. KG
carl-zeiss strasse 7-9
saalfeld 07318
GM  07318
Manufacturer Contact
steffen ulbrich
carl-zeiss strasse 7-9
saalfeld 07318
GM   07318
MDR Report Key12279927
MDR Text Key265155450
Report Number9681407-2021-00021
Device Sequence Number1
Product Code FSY
Combination Product (y/n)N
Reporter Country CodeAS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial
Report Date 08/03/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/05/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model Number4065520
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received08/03/2021
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-