Model Number 10-80-00 |
Device Problem
Decreased Pump Speed (1500)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 07/09/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
There was no patient involvement.Serial number is unknown.This information will be provided in a supplemental report if made available.As the serial number is unknown, the device manufacture date could not be determined.This information will be provided in a supplemental report if made available.Livanova (b)(4) manufactures the s5 roller pump.The incident occurred in (b)(6).A livanova field service representative was dispatched to the facility to investigate the device and could not confirm the reported issue.The encoder and the processor board were replaced as precaution.Subsequent functional verification testing was completed without further issues and the unit was returned to service.A video was shared with livanova.The results of the analysis will be provided in a supplemental report.
|
|
Event Description
|
Livanova received report that the rpms on a s5 roller pump dropped to zero twice during setup.There was no patient involvement.
|
|
Manufacturer Narrative
|
H.10: the most likely root cause is a temporary communication issue between the processor board and the shaft angle encoder.
|
|
Event Description
|
See initial report.
|
|