• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS ROCHE DIAGNOSTICS COBAS ELECSYS ANTI-TPO; SYSTEM, TEST, THYROID AUTOANTIBODY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS ROCHE DIAGNOSTICS COBAS ELECSYS ANTI-TPO; SYSTEM, TEST, THYROID AUTOANTIBODY Back to Search Results
Catalog Number 07026935190
Device Problem High Test Results (2457)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/21/2021
Event Type  malfunction  
Manufacturer Narrative
The customer declined to provide any additional information for the investigation.From the information provided, a general reagent issue can be excluded.The cause of the event could not be determined.The event is consistent with a pre-analytical issue (micro clots or fibrin filaments in the sample) and/or the presence of foam/bubbles on the reagent surface.The investigation did not identify a product problem.
 
Event Description
The initial reporter complained of discrepant results for 1 patient sample tested for elecsys anti-tpo (anti-tpo) on a cobas e801 module.The initial result was 51.7 iu/ml.The repeat was < 9 iu/ml.No questionable results were reported outside of the laboratory.The e801 module serial number was not provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROCHE DIAGNOSTICS COBAS ELECSYS ANTI-TPO
Type of Device
SYSTEM, TEST, THYROID AUTOANTIBODY
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
michael leslie
9115 hague road
na
indianapolis, IN 46250
3175214343
MDR Report Key12281722
MDR Text Key268317602
Report Number1823260-2021-02289
Device Sequence Number1
Product Code JZO
UDI-Device Identifier04015630939626
UDI-Public04015630939626
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K051890
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 08/05/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/05/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number07026935190
Device Lot Number526599
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/21/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-