The customer contacted a siemens customer care center (ccc).Quality controls (qc) recovered in range at the time of the event.Only one patient sample was affected.The customer followed the guidance in the n latex flc lambda instructions for use (ifu) and repeated the results in a higher sample dilution.According to the n latex flc lambda ifu, the following information should be considered: "excessively elevated monoclonal immunoglobulin concentrations might have the potential to suppress the reaction of anti-flc antibodies with free light chain molecules.If results do not fit to previous ones, or to results of other tests (e.G.Serum immune electrophoresis, immunofixation, differential blood cell count) and/or to the clinical situation, it is recommended to re-analyze the sample in a higher sample dilution." the cause of the event is unknown.The reagent is performing according to specifications.No further evaluation of this device is required.
|
A discordant, falsely low free light chains, type lambda (flc lambda) result was obtained on a patient sample using a 1:20 dilution on a bn ii system using n latex flc lambda reagent.The result was not reported to the physician(s).The sample was then repeated for flc lambda again using a 1:20 dilution on the same system using the same reagent, recovering higher.This result was not reported to the physician(s).The sample was then repeated for flc lambda using a 1:100 dilution on the same system using the same reagent, recovering higher.This result was not reported to the physician(s).The sample was then repeated for flc lambda using a 1:400 dilution on the same system using the same reagent, recovering higher.This result was reported, as the correct result, to the physician(s).The sample was then repeated for flc lambda using a 1:2000 dilution, recovering higher.This result was not reported to the physician(s).There are no known reports of patient intervention or adverse health consequences due to the discordant free light chains, type lambda (flc lambda) results.
|