• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL ADVISOR HD GRID MAPPING CATHETER, SENSOR ENABLED; CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL ADVISOR HD GRID MAPPING CATHETER, SENSOR ENABLED; CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING Back to Search Results
Model Number D-AVHD-DF16
Device Problem Entrapment of Device (1212)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/28/2021
Event Type  malfunction  
Event Description
During the procedure, the catheter could not be advanced through the sheath and was lodged in the groin.The catheter was cut and a larger introducer was used to remove the catheter.The procedure was completed without replacing the device with no consequences to the patient.
 
Manufacturer Narrative
One bi-directional, curve d-f, sensor enabled, advisor hd grid mapping catheter was received for evaluation.The electrodes were displaced and the pellethane tubing and splines were corrugated and torn.The paddle and shaft was returned cut from the catheter handle.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the torn tubing, cut shaft, displaced electrodes and insertion difficulty is consistent with damage during use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVISOR HD GRID MAPPING CATHETER, SENSOR ENABLED
Type of Device
CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING
Manufacturer (Section D)
ST. JUDE MEDICAL
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS  1897-4050
MDR Report Key12299907
MDR Text Key265782789
Report Number3008452825-2021-00405
Device Sequence Number1
Product Code DRF
UDI-Device Identifier05415067028198
UDI-Public05415067028198
Combination Product (y/n)N
PMA/PMN Number
K172393
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 08/17/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/10/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/31/2024
Device Model NumberD-AVHD-DF16
Device Catalogue NumberD-AVHD-DF16
Device Lot Number7944975
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/05/2021
Was the Report Sent to FDA? No
Date Manufacturer Received08/17/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-