STRYKER ENDOSCOPY-SAN JOSE PACKAGING, FORMULA 180 SHAVER (HAND CONTROL); BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL
|
Back to Search Results |
|
Model Number 0375708500 |
Device Problems
Melted (1385); Adverse Event Without Identified Device or Use Problem (2993)
|
Patient Problems
Superficial (First Degree) Burn (2685); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 07/13/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
|
|
Event Description
|
It was reported that there was a thermal event.
|
|
Event Description
|
It was reported that there was a thermal event.
|
|
Manufacturer Narrative
|
The product was returned for investigation and the reported failure mode was not confirmed.Alleged failure: the bur it was over spinning nad melted the plastic.The probable root cause/s could be the handpiece was corrupt due to (third-party components- power cable), misuse, the suction valve was not activated.The device manufacture date is not known.The failure mode will be monitored for future reoccurrence.
|
|