• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ENDOSCOPY-SAN JOSE PACKAGING, FORMULA 180 SHAVER (HAND CONTROL); BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ENDOSCOPY-SAN JOSE PACKAGING, FORMULA 180 SHAVER (HAND CONTROL); BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL Back to Search Results
Model Number 0375708500
Device Problems Melted (1385); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Superficial (First Degree) Burn (2685); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/13/2021
Event Type  malfunction  
Manufacturer Narrative
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
 
Event Description
It was reported that there was a thermal event.
 
Event Description
It was reported that there was a thermal event.
 
Manufacturer Narrative
The product was returned for investigation and the reported failure mode was not confirmed.Alleged failure: the bur it was over spinning nad melted the plastic.The probable root cause/s could be the handpiece was corrupt due to (third-party components- power cable), misuse, the suction valve was not activated.The device manufacture date is not known.The failure mode will be monitored for future reoccurrence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PACKAGING, FORMULA 180 SHAVER (HAND CONTROL)
Type of Device
BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
MDR Report Key12304967
MDR Text Key267531889
Report Number0002936485-2021-00419
Device Sequence Number1
Product Code GFA
UDI-Device Identifier07613327057645
UDI-Public07613327057645
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Type of Report Initial,Followup
Report Date 10/03/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/10/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0375708500
Device Catalogue Number0375708500
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/11/2021
Date Manufacturer Received07/13/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-