• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER ORTHOPAEDIC MFG. LTD. PERSONA ARTICULAR SURFACE MEDIAL CONGRUENT (MC) LEFT 11 MM HEIGHT; PROSTHESIS, KNEE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER ORTHOPAEDIC MFG. LTD. PERSONA ARTICULAR SURFACE MEDIAL CONGRUENT (MC) LEFT 11 MM HEIGHT; PROSTHESIS, KNEE Back to Search Results
Model Number N/A
Device Problem Difficult to Insert (1316)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/19/2021
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Report source - foreign: (b)(6).The complainant has indicated that the product will be returned to zimmer biomet for investigation.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported that total knee arthroplasty was performed.During the procedure, the surgeon tried to insert the articular surface to the tibia plate.However, the articular surface didn't mate with the tibia plate.Several insertions were done, but the insertions failed.Subsequently, another articular surface was used to complete the procedure.Attempts have been made, but no further information available.
 
Event Description
No further event information available at the time of this report.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Visual inspection of the returned device identified that the dovetail feature was flared out.The device history records were reviewed and no discrepancies were identified.A definitive root cause could not be determined with the information provided.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERSONA ARTICULAR SURFACE MEDIAL CONGRUENT (MC) LEFT 11 MM HEIGHT
Type of Device
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER ORTHOPAEDIC MFG. LTD.
building no 2 east park
shannon industrial estate
shannon, county clare
EI 
MDR Report Key12313489
MDR Text Key266223786
Report Number3007963827-2021-00172
Device Sequence Number1
Product Code MBH
UDI-Device Identifier00889024468986
UDI-Public(01)00889024468986(17)260208(10)64953207
Combination Product (y/n)N
PMA/PMN Number
K150090
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 10/14/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/12/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number42512100411
Device Lot Number64953207
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/13/2021
Was the Report Sent to FDA? No
Date Manufacturer Received10/13/2021
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
-
-