Age at time of event: 18 years or older.The returned product consisted of the rotapro atherectomy system.The burr catheter was received attached to the advancer unit.The rotawire used in the procedure was also returned.The advancer, handshake connections, sheath, coil, burr and annulus were visually and microscopically examined.Inspection presented no damage or irregularities to the device.Functional testing was performed by using a test wireclip torquer.During testing, the returned rotawire was able to be inserted into the wireclip torquer, and the wireclip torquer was able to be inserted into the rotapro advancer.The brake defeat button was held down with the wireclip torquer inserted with no issues.Functional testing was performed by attempting to rotate the drive shaft, and the drive shaft was not able to be rotated.The rotapro advancer was then connected to the rotapro control console system.The knob switch (ablation button) was pressed, but the device stalled and would not get any speed.In order to determine the cause of the stall, destructive testing was performed.The advancer was dismantled and the components inside the advancer were inspected, but destructive testing was not conclusive as there were no damages to the components.There is not enough evidence to definitively say what the cause of the device stall was.Product analysis could not confirm the reported events, as the wireclip torquer was able to be inserted and held down the brake defeat button with no issues.The reported advancer activation issues could not be confirmed, as the device stalled when the ablation button was pressed.Therefore, it was not determined if the knob switch could turn the advancer on and off.
|