SIEMENS HEALTHCARE GMBH- AT ARTIS ZEE CEILING; INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM Back to Search Results
Model Number 10094137
Device Problem Mechanical Problem (1384)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/11/2021
Event Type  malfunction  
Manufacturer Narrative
Siemens is conducting a thorough investigation of the reported events.As this event is under investigation, a root cause has not yet been determined.A supplement report will be filed upon completion of the investigation.
 
Event Description
It was reported to siemens that a malfunction occurred while operating the artis zee ceiling system.The user reported that the accessory rail on the table is missing a stop and it dropped on the operators foot.The operator did not require medical intervention.Siemens has requested additional information in order to conduct an investigation of the reported event.
 
Manufacturer Narrative
Siemens has completed an investigation of the reported event.The root cause was determined to be mechanical damage.The system in question is equipped with a special metal rail on the side of the patient table to which various accessories and options can be attached.In this case, the metal rail was mechanically damaged and unstable to such an extent that it apparently fell off.Under normal circumstances, this cannot happen because the rail has a stop that prevents this.It was no longer possible to determine beyond doubt who or what caused this damage to the system.However, experts agree that external force (e.G., improper use) are the most likely causes.A possible accumulation of faults or even a possible general fault that would require corrective measures of the installed base could not be determined by the investigation.After detailed investigation, the incident is not classified as a reportable event as neither serious injury, death nor an unexpected, prolonged hospitalization of the patient or any other person occurred or could be expected.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARTIS ZEE CEILING
Type of Device
INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM
Manufacturer (Section D)
SIEMENS HEALTHCARE GMBH- AT
siemensstrasse 1
forchheim, germany 91301
GM  91301
MDR Report Key12341356
MDR Text Key267232824
Report Number3004977335-2021-92856
Device Sequence Number1
Product Code OWB
UDI-Device Identifier04056869010052
UDI-Public04056869010052
Combination Product (y/n)N
PMA/PMN Number
K181407
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Remedial Action Repair
Type of Report Initial,Followup
Report Date 08/19/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/19/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number10094137
Was Device Available for Evaluation? Yes
Date Manufacturer Received10/04/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-