Model Number 10374 |
Device Problem
Leak/Splash (1354)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/09/2021 |
Event Type
malfunction
|
Event Description
|
It was reported there was a leak in the hemostatic valve.A left atrial appendage (laa) closure procedure was being performed.A watchman access system (was) was positioned and a watchman laa closure device & delivery system (wds) were used.During the procedure the hemostatic valve could not be tightened enough.As a result of this there was blood leaking from the hemostatic valve while the devices were in the patient.There were no complications to the patient.The procedure was completed successfully with another was and the closure device was implanted in the laa of the patient.
|
|
Event Description
|
It was reported there was a leak in the hemostatic valve.A left atrial appendage (laa) closure procedure was being performed.A watchman access system (was) was positioned and a watchman laa closure device & delivery system (wds) were used.During the procedure the hemostatic valve could not be tightened enough.As a result of this there was blood leaking from the hemostatic valve while the devices were in the patient.There were no complications to the patient.The procedure was completed successfully with another was and the closure device was implanted in the laa of the patient.
|
|
Manufacturer Narrative
|
Device/media analysis: the returned product consisted of a watchman access system (was) with the dilator in the sheath.The device was microscopically and visually examined.There was damage to the silicone valve.The device was functionally tested by attaching a syringe (filled with water), and positive/negative pressure was applied with the valve open and closed.There was a leak at the valve.Inspection of the remainder of the device presented no other damage or irregularities.Product analysis confirmed the reported leak.
|
|
Search Alerts/Recalls
|