• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST PAUL LEVEL 1 TRAUMA FAST FLOW SYSTEMS; WARMER, BLOOD, NON-ELECTROMAGNETIC RADIATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST PAUL LEVEL 1 TRAUMA FAST FLOW SYSTEMS; WARMER, BLOOD, NON-ELECTROMAGNETIC RADIATION Back to Search Results
Model Number 8002950
Device Problems Display or Visual Feedback Problem (1184); Failure to Pump (1502)
Patient Problems Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
It was reported that the device was not pumping or warming and sometimes the display was not showing.The issue was discovered during testing.
 
Manufacturer Narrative
Other text: no causes or potential causes of the customer's reported problem were found during the review of service and repair records.A product sample was received for evaluation.Visual and functional testing were performed.Visual inspection found device in good condition with no physical damage.Functional testing found device was not recirculating water.Removed back panel for further investigation.Visual inspection and found that there was a crack in the return tube which had leaked water, and damaged multiple internal components.Replaced main printed circuit board (pcb), return tube, and membrane switch.Device passed all functional and electrical tests.The root cause of the reported issue was found to be a design issue.An internal capa has been opened and this issue will be monitored for any increase in occurrence.If the occurrence of this issue increases significantly, additional corrective actions will be taken.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL 1 TRAUMA FAST FLOW SYSTEMS
Type of Device
WARMER, BLOOD, NON-ELECTROMAGNETIC RADIATION
Manufacturer (Section D)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer (Section G)
NULL
1265 grey fox rd.
st. paul MN 55112
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key12342675
MDR Text Key267277770
Report Number3012307300-2021-08608
Device Sequence Number1
Product Code BSB
UDI-Device Identifier50695085829506
UDI-Public50695085829506
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
BK020043
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 03/29/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/19/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number8002950
Device Catalogue Number8002950
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/28/2021
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received12/19/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/05/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-