U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. OUTER-GRIP LOCKING FEM IMPLANT IMPACTOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. OUTER-GRIP LOCKING FEM IMPLANT IMPACTOR Back to Search Results
Model Number 74018902
Device Problem Scratched Material (3020)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/04/2021
Event Type  malfunction  
Event Description
It was reported that during set up for a tka that the tip of an outer-grip locking fem implant impactor was found to be chipped.No delay reported, procedure was able to be performed with a backup device.No patient harm reported.
 
Manufacturer Narrative
H3, h6: the associated device, intended for use in treatment, was returned and evaluated.A visual inspection of the returned device confirmed the stated failure mode.The tips of the device have broken off, rendering the device inoperative.The device shows signs of extensive use.A review of complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OUTER-GRIP LOCKING FEM IMPLANT IMPACTOR
Type of Device
IMPACTOR
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key12342749
MDR Text Key267279366
Report Number1020279-2021-06467
Device Sequence Number1
Product Code HWA
UDI-Device Identifier00885556697962
UDI-Public00885556697962
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 08/25/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/19/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number74018902
Device Catalogue Number74018902
Device Lot Number21AM15754
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/20/2021
Date Manufacturer Received08/25/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-