Model Number N/A
Device Problems Positioning Failure (1158); Difficult to Insert (1316)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/28/2021
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Concomitant medical products: associated product tibia cemented 5 degree stemmed right size e item# 42532007102, lot# 64721507.Report source: (b)(6).Customer has indicated that the product will be returned to zimmer biomet for investigation.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported that during a knee procedure, the surgeon failed to insert an articular surface to the tibial component.Several attempts were made and he confirmed that there was no cement or third particles between the articular surface and tibial component.Subsequently, a backup was used to complete the procedure.Attempts to obtain additional information have been made; however, no more information is available.
 
Manufacturer Narrative
This follow-up is being submitted to relay additional information.Visual inspection of the returned articular surface found that the dovetail feature was flared out.Medical records were not provided.Device history record (dhr) was reviewed for deviations and/ or anomalies with no deviations / anomalies identified.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARTICULAR SURFACE
Type of Device
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER ORTHOPAEDIC MFG. LTD.
building no 2 east park
shannon industrial estate
shannon, county clare
EI 
MDR Report Key12348470
MDR Text Key267472249
Report Number3007963827-2021-00176
Device Sequence Number1
Product Code MBH
UDI-Device Identifier00889024243750
UDI-Public(01)00889024243750(17)250131(10)64549154
Combination Product (y/n)N
PMA/PMN Number
K121771
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 09/28/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/20/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number42522400710
Device Lot Number64549154
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/08/2021
Was the Report Sent to FDA? No
Date Manufacturer Received09/22/2021
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
Treatment
SEE H10.
-
-