MAUDE Adverse Event Report: UNKNOWN ROTAREX SET 6F 135CM; THROMBECTOMY & ATHERECTOMY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

UNKNOWN ROTAREX SET 6F 135CM; THROMBECTOMY & ATHERECTOMY Back to Search Results
Catalog Number 80237
Device Problem Stretched (1601)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/26/2021
Event Type  malfunction  
Manufacturer Narrative
As the lot number for the device was provided, a review of the device history records is currently being performed.The return of the sample is pending.The investigation of the reported event is currently underway.The medical device manufacturer and manufacturing location for the straub product was selected as unknown due to system limitations.The correct medical device manufacturer and manufacturing location are straub medical us.(expiry date: 07/2023).
 
Event Description
It was reported that during a procedure, the guidewire of the device allegedly started to unwind upon removal.There was no reported patient injury.
 
Manufacturer Narrative
Manufacturing review: a manufacturing review was conducted and there was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: the sample was not returned for evaluation.Therefore, the investigation is inconclusive for the reported issue.The definitive root cause could not be determined based upon available information.Labeling review: a review of product labeling documentation (e.G.Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.The medical device manufacturer (d3) and manufacturing location (g1) for the straub product was selected as unknown due to system limitations.The correct medical device manufacturer and manufacturing location are straub medical us.D4 (expiry date: 07/2023).H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device not returned.
 
Event Description
It was reported that during a procedure, the guidewire of the device allegedly started to unwind upon removal.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROTAREX SET 6F 135CM
Type of Device
THROMBECTOMY & ATHERECTOMY
Manufacturer (Section D)
UNKNOWN
BR 
Manufacturer (Section G)
UNKNOWN
BR  
Manufacturer Contact
judy ludwig
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key12352009
MDR Text Key267626836
Report Number3008439199-2021-00132
Device Sequence Number1
Product Code MCW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K172315
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/23/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/22/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number80237
Device Lot Number201041
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/23/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-