DORNOCH DUO FLUID CART; APPARATUS, SUCTION, WARD USE, PORTABLE, AC-POWERED Back to Search Results
Model Number N/A
Device Problems Overheating of Device (1437); Electrical Power Problem (2925)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/21/2021
Event Type  malfunction  
Manufacturer Narrative
This event has been recorded by zimmer biomet under (b)(4).Review of the most recent repair record determined that the device was not powering on from wall power and that the power inlet module and cord were burnt.The power inlet module was replaced and resolved the reported issue.It was noted during testing that the unit had low suction and the inline filter and vacuum regulator were replaced.Device is used for treatment.A definitive root cause cannot be determined.The event is confirmed.
 
Event Description
It was reported this unit will not power on via wall power.Found it will power on via the evacuation unit.The event timing was during cleaning.Investigation found the power inlet module was burnt and the cord had heat damage.No adverse event was reported as a result of this malfunction.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DUO FLUID CART
Type of Device
APPARATUS, SUCTION, WARD USE, PORTABLE, AC-POWERED
Manufacturer (Section D)
DORNOCH
200 northwest parkway
riverside MO 64150
Manufacturer (Section G)
DORNOCH
200 northwest parkway
riverside MO 64150
Manufacturer Contact
christina arnt
56 e. bell drive
warsaw, IN 46582
5745273773
MDR Report Key12352998
MDR Text Key267645957
Report Number0001954182-2021-00051
Device Sequence Number1
Product Code JCX
UDI-Device Identifier00889024465992
UDI-Public(01)00889024465992
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K190789
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 08/23/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/23/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number00514010100
Device Lot Number0042750
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received08/20/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/02/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
-
-