Model Number IPN040620 |
Device Problem
Connection Problem (2900)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 07/30/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Complaint verification testing could not be performed as no sample was returned for analysis.A device history record review could not be conducted since the lot number of the device was not provided.Without the device to evaluate, the complaint could not be confirmed and the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.If the sample becomes available at a later date a follow up report will be submitted with investigation results.
|
|
Event Description
|
It was reported that "the connection part between the nut adaptor and the flowmeter was so unstable that the adaptor got detached during use.Therefore, it was replaced with a new unit".No patient harm or injury reported.Patient condition reported as "fine".
|
|
Event Description
|
It was reported that "the connection part between the nut adaptor and the flowmeter was so unstable that the adaptor got detached during use.Therefore, it was replaced with a new unit".No patient harm or injury reported.Patient condition reported as "fine".
|
|
Manufacturer Narrative
|
(b)(4).The sample was returned for evaluation.A visual exam was performed and it was observed that the puncture pin was missing.No other issues were observed.A puncture pin was obtained and oxygen entrainment testing was performed.During the setup it was observed that the assembly of the nut adaptor and the upper body was unstable.Even with that condition, the sample was able to be tested on oxygen entrainment testing, but the sample failed the test.A visual inspection was performed and wear was found on the internal tabs of the adaptor.Based on the investigation performed, the complaint is confirmed.Although the complaint is confirmed, there is not sufficient evidence to assure that this issue was originated during the manufacturing assembly or molding process.The wear on the internal tabs of the adaptor was most likely caused by the end user during the connection of the adaptor into the flowmeter that causes an unstable connection issue.
|
|