MAUDE Adverse Event Report: HOYA CORPORATION PENTAX TOKYO OFFICE PENTAX; IMAGINA GASTROSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HOYA CORPORATION PENTAX TOKYO OFFICE PENTAX; IMAGINA GASTROSCOPE Back to Search Results
Model Number EG29-I10C
Device Problems Moisture Damage (1405); Optical Distortion (3000)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/15/2021
Event Type  malfunction  
Manufacturer Narrative
This device is not distributed in us so that unique identifier is blank.This device is not distributed in the usa, therefore the pma/510(k) number is not applicable.Evaluation summary: we checked the returned product and confirmed that a fluid damage on the cmos module due to lg cable connector assy leak.This report is being filed as part of the pentax backlog management plan.
 
Event Description
Foggy image.Led moisture.Blue marking at cmos-cable, no leak detected.There was no report of patient harm.The time of event is unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENTAX
Type of Device
IMAGINA GASTROSCOPE
Manufacturer (Section D)
HOYA CORPORATION PENTAX TOKYO OFFICE
tsutsujigaoka 1-1-110
akishima-shi, tokyo 196-0 012
JA  196-0012
Manufacturer (Section G)
HOYA CORPORATION PENTAX TOKYO OFFICE
tsutsujigaoka 1-1-110
akishima-shi, tokyo 196-0 012
JA   196-0012
Manufacturer Contact
william goeller
3 paragon drive
montvale, NJ 07645
8004315880
MDR Report Key12357373
MDR Text Key267857918
Report Number9610877-2021-00683
Device Sequence Number1
Product Code FDS
Combination Product (y/n)N
Reporter Country CodeGM
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Physician
Type of Report Initial
Report Date 08/24/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/24/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberEG29-I10C
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/15/2021
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-