• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD PERIPHERAL VASCULAR, INC. ENCOR MRIPROBE TROCAR10G; BIOPSY INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD PERIPHERAL VASCULAR, INC. ENCOR MRIPROBE TROCAR10G; BIOPSY INSTRUMENT Back to Search Results
Model Number ECPMR0110G
Device Problem Device Contamination with Chemical or Other Material (2944)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/28/2021
Event Type  malfunction  
Manufacturer Narrative
As the lot number for the device was not provided, a review of the device history records could not be performed.The device has been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.
 
Event Description
It was reported that during a biopsy procedure, the device allegedly had foreign material on sample cup.It was further reported that the foreign material was removed and the needle was used.There was no reported patient injury.
 
Manufacturer Narrative
H10: manufacturing review: the lot number was not provided; however, the manufacturing process was reviewed.Based on the results of the manufacturing review, it was concluded that contamination on the complaint products could be related to the nature of parts and manufacturing process of probe products.The products have been 100% inspected to remove contamination however, additional operator training was conducted to ensure awareness and focusing on the product contamination in their operations.Investigation summary: the physical device was not returned for evaluation.One biohazard bag was received with an unknown substance/material.The material was inspected via microscopic imaging and appeared to be amber in color with a jelly like consistency, containing fibers within along with what appears to be blue flecks.The material was further inspected via ftir/sem analysis.The analysis concluded the returned material composition was consistent with pe (polyethylene, pet (polyethylene terephthalate), silicone, dispersant (such as drewsperse s-825), and 300 series stainless (such as 304 stainless) with some aluminum (ai) detected which may be the result of metal transfer as the material is not consistent with stainless.Although the ftir/sem analysis concluded material composition consistent with materials used in the build of this device, neither the physical device nor a photo identifying the alleged foreign material were provided; therefore the investigation is inconclusive.The returned material could potentially be a result of handling and usage by the end user or related to the nature of parts and manufacturing process of probe products.Based on the provided information, the foreign material like fuzz was found and removed from the device and unknown if the foreign material was potentially contaminated with any other materials prior to evaluation.A definitive root cause for the reported issue could not be determined.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device not returned.
 
Event Description
It was reported that during preparation for a biopsy procedure, foreign material like fuzz was allegedly found on sample cup.It was further reported that the foreign material was removed and the needle was used.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENCOR MRIPROBE TROCAR10G
Type of Device
BIOPSY INSTRUMENT
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
Manufacturer (Section G)
INFUS MEDICAL (THAILAND)
706 moo 4
bangpoo ind estate
samutprakarn province 10280
TH   10280
Manufacturer Contact
judy ludwig
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key12357440
MDR Text Key267806731
Report Number2020394-2021-01494
Device Sequence Number1
Product Code KNW
UDI-Device Identifier00801741086304
UDI-Public(01)00801741086304
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K040842
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 03/03/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/24/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberECPMR0110G
Device Catalogue NumberECPMR0110G
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Date Returned to Manufacturer08/19/2021
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/11/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-