The device was returned and evaluated.It was confirmed that the microtip needle had wear on its outer surface near the adjacent, distal end of the needle sheath.The tip of the needle that contacts tissue was also bent over, consistent with forceful use on hard tissue, such as bone, which is outside the indications for use.The user also did report using the microtip for ankle debridement.Such use could apply a moment to the needle tip and push it against the inside of the needle sheath, causing friction and wear between the 2 parts and the generation of metal debris.
|
During a procedure with the tenex system, metal fragments were generated near the tip of the needle and ejected into the treatment site.The procedure was stopped, and the fragments removed with a separate instrument.There was a 5-10 delay due to the additional action.There were no patient complications.
|