No product was returned as the involved device remains in situ and no information could be provided confirming the type of infection or its severity.The issue was discovered via a literature review (evaluation of the effectiveness of expandable cages for reconstruction of the anterior column of the spine) and no additional information could be obtained.All nuvasive instrumentation is cleaned and sterilized at the user facility pre case, no sterilization records could be provided.The root cause of the infection could not be confirmed or what the infection was but it is believed to be related to environmental contamination as there was no reported association with the device that currently remains in patient.".Potential adverse events and complications.As with any major surgical procedures, there are risks involved in orthopedic surgery.Infrequent operative and postoperative complications that may result in the need for additional surgeries include: early or late infection.Potential risks identified with the use of this system, which may require additional surgery, include.Infection." ".Cleaning and decontamination.All non-sterile instruments must first be thoroughly cleaned using the validated methods prescribed in the nuvasive cleaning and sterilization instructions (doc #9400896) before sterilization and introduction into a sterile surgical field.Contaminated instruments should be wiped clean of visible soil at the point of use, prior to transfer to a central processing unit for cleaning and sterilization.The validated cleaning methods include both manual and automated cleaning.Visually inspect the instruments following performance of the cleaning instructions to ensure there is no visual contamination of the instruments prior to proceeding with sterilization.If possible contamination is present at visual inspection, repeat the cleaning steps.Contaminated instruments should not be used, and should be returned to nuvasive.Contact your local representative or nuvasive directly for any additional information related to cleaning of nuvasive surgical instruments." ".Sterilization.All non-sterile instruments and implants are sterilizable by steam autoclave using standard hospital practices, in addition to nuvasives validated parameters.In a properly functioning and calibrated steam sterilizer, effective sterilization may be achieved using the parameters prescribed in the nuvasive cleaning and sterilization instructions (doc # (b)(4))&".
|