ANGIOMED GMBH & CO. MEDIZINTECHNIK KG COVERA VASCULAR COVERED STENT Back to Search Results
Model Number AVSM08040
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Ischemia (1942)
Event Date 05/15/2019
Event Type  Injury  
Manufacturer Narrative
As the lot number for the device was provided, a review of the device history records will be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.(expiry date: 03/2017).Device not returned.
 
Event Description
It was reported through the results of a clinical trial, that approximately two years and five months post index procedure, new lesion stenosis was observed and covera vascular covered stent was used to successfully treat the target lesion.The patient was diagnosed with right hand ischemia related to study procedure and device.The patient current status is unknown.
 
Event Description
It was reported through the results of a clinical trial, that approximately two years and five months post index procedure, new lesion stenosis was observed and covera vascular covered stent was used to successfully treat the target lesion.The patient was diagnosed with right hand ischemia related to study procedure and device.The patient current status is unknown.
 
Manufacturer Narrative
H10: manufacturing review: a review of manufacturing records was not performed, as additional complaints have not been reported for this lot.Based on the information available it is not reasonably suggested that a manufacturing process may have caused or contributed to the reported issue.However, the lot history records of this lot were reviewed with special attention to the manufacturing and inspection of this product and the product was found to have met the specification prior to shipment.Investigation summary: the stent was not returned for evaluation as it remains implanted.No x-ray images were provided for evaluation.It was noted that stenosis occurred in the access circuit, pre dilation as well as post dilation was performed; there was no report of any irregularities or complication during initial stent placement.Based on the information available the investigation is closed with inconclusive result.A definite root cause for the reported event could not be determined.Labeling review: in reviewing the relevant labeling for this product, it was found that the instructions for use sufficiently address the potential risk.Based on the instructions for use complications and adverse events associated with the use of the covera plus vascular covered stent may include the usual complications associated with endovascular stent and covered stent placement and dialysis shunt revisions, which includes restenosis of the target vessel.Regarding pre and post dilation the instructions for use states: "pre-dilate the stenosis with a pta balloon catheter of appropriate length and diameter for the lesion to be treated" and "post dilate the covered stent with an angioplasty balloon sized appropriately as to ensure complete wall apposition to the reference vessel.Avoid balloon dilation in the healthy, non-stenosed segment of the vessel." h11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COVERA VASCULAR COVERED STENT
Type of Device
VASCULAR COVERED STENT
Manufacturer (Section D)
ANGIOMED GMBH & CO. MEDIZINTECHNIK KG
wachhausstr. 6
karlsruhe 76227
GM  76227
Manufacturer (Section G)
ANGIOMED GMBH & CO. MEDIZINTECHNIK KG
wachhausstr. 6
karlsruhe 76227
GM   76227
Manufacturer Contact
judy ludwig
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key12369763
MDR Text Key268205726
Report Number9681442-2021-00461
Device Sequence Number1
Product Code PFV
UDI-Device Identifier00801741106286
UDI-Public(01)00801741106286
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P170042
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Study,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/18/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/26/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberAVSM08040
Device Catalogue NumberAVSM08040
Device Lot NumberANAQ2171
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/18/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
ASPIRIN, APIXABAN, CILOSTAZOL; ASPIRIN, APIXABAN, CILOSTAZOL; ASPIRIN, APIXABAN, CILOSTAZOL
Patient Outcome(s) Other;
Patient Age74 YR
Patient SexFemale
Patient Weight60 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-