• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET CARDIOPULMONARY GMBH HLS SET ADVANCED 7.0; OXYGENATOR, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET CARDIOPULMONARY GMBH HLS SET ADVANCED 7.0; OXYGENATOR, CARDIOPULMONARY BYPASS Back to Search Results
Model Number HLS SET
Device Problem Loose or Intermittent Connection (1371)
Patient Problem Insufficient Information (4580)
Event Type  malfunction  
Manufacturer Narrative
The investigation is ongoing.A follow up medwatch will be submitted when additional information becomes available.
 
Event Description
It was reported that the hls module was disconnected from the cardio help during patient treatment.Complaint id # (b)(4).
 
Manufacturer Narrative
It was reported that the hls module was disconnected from the cardiohelp during patient treatment.No harm to any person reported.They changed the set out and tried to replicate the error with the old hls set using a second cardiohelp unit and could not replicate.The affected product was not available for further investigation as it was discarded by the customer without any notes of the lot number.No visible damage was noticed on the hls set or the cardiohelp, neither before nor after the event by the customer.Based on the given information, the cause of the event could not be attributed to the device.Thus the reported failure could not be confirmed.According to the hls risk assessment (dms#1468452, v26) the most probable root cause is that hls module was not securely fixated into the cardiohelp-i-drive.In the instruction for use ( hls set advanced 5.0 / 7.0, hit set advanced 5.0 / 7.0, v1.5, chapter 5.3.1 safety instructions for the oxygenator) is stated to ensure that the device is fitted onto the drive correctly and securely fixed, to eliminate the risk of magnetic decoupling between the drive and the centrifugal pump.In order to avoid the reoccurrence of the reported event the ssu (sales and service unit) should inform the user about the investigation results and the above mentioned sections of the ifu.The occurrence rate was calculated for the reported issue and it was determined that this is not a systemic issue.Therefore, no remedial action is required.The occurrence rate related to the reported issue is currently being monitored as part of maquet cardiopulmonary¿ s trending program and additional investigations or corrections will be implemented in case of adverse trending.
 
Event Description
Complaint id #(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HLS SET ADVANCED 7.0
Type of Device
OXYGENATOR, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
MAQUET CARDIOPULMONARY GMBH
neue rottenburger strasse 37
hechingen
Manufacturer (Section G)
JULIA KAPFENBERGER
neue rottenburger strasse 37
hechingen
Manufacturer Contact
neue rottenburger strasse 37
hechingen 
MDR Report Key12376469
MDR Text Key268425673
Report Number8010762-2021-00476
Device Sequence Number1
Product Code DTZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K112360
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Consumer,Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 01/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/27/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberHLS SET
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/05/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-