• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS WINTER & IBE GMBH HF-RESECTION ELECTRODE, LOOP, 24 FR., 0.2 WIRE, MEDIUM, 30, STERILE, SINGLE USE; HF-RESECTION ELECTRODES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS WINTER & IBE GMBH HF-RESECTION ELECTRODE, LOOP, 24 FR., 0.2 WIRE, MEDIUM, 30, STERILE, SINGLE USE; HF-RESECTION ELECTRODES Back to Search Results
Model Number WA22306D
Device Problems Break (1069); Fracture (1260); Material Fragmentation (1261); Mechanical Problem (1384)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  Injury  
Manufacturer Narrative
The suspect medical device has not yet been returned to olympus for evaluation/investigation.Therefore, the exact cause of the patient¿s outcome and the reported phenomenon could not be determined and is being judged as unknown.However, if the suspect medical device is returned for evaluation/investigation or additional significant information becomes available, this report will be updated.
 
Event Description
Olympus was informed that during an unspecified procedure, the loop wire at the distal end of two hf resection electrodes of the same lot broke off and fragments fell into the patient.It was only possible to locate and remove the broken fragments by using an ultrasound device.However, this led to complications and resulted in a significant prolongation of the procedural time.There are no reports of any complication after the surgery.
 
Manufacturer Narrative
The suspect medical devices were not returned to the manufacturer for evaluation/investigation since they were reportedly discarded by the user facility.Therefore, the exact cause of the patient's outcome and the reported phenomenon could not be conclusively determined and is being judged as unknown.However, the reported failure is a known phenomenon and can very likely be attributed to component failure caused by use-related wear and tear.A material or quality problem can be excluded since a manufacturing and quality control review was performed for the affected lot number of the resection electrodes without showing any abnormalities.The case will be closed on olympus side with no further actions.The reported event/incident will be recorded for trending and surveillance purposes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HF-RESECTION ELECTRODE, LOOP, 24 FR., 0.2 WIRE, MEDIUM, 30, STERILE, SINGLE USE
Type of Device
HF-RESECTION ELECTRODES
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg 22045
GM  22045
MDR Report Key12376616
MDR Text Key268433267
Report Number9610773-2021-00235
Device Sequence Number1
Product Code FAS
UDI-Device Identifier14042761051689
UDI-Public14042761051689
Combination Product (y/n)N
PMA/PMN Number
K100275
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Type of Report Initial,Followup
Report Date 09/21/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/27/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberWA22306D
Device Catalogue NumberWA22306D
Device Lot Number1000063179
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received09/17/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-