Reportable based on device analysis completed on 06aug2021.It was reported that there was difficulty advancing the device over the wire.A 2.1mm jetstream xc catheter was selected for a patient procedure for a right superficial femoral artery occlusion.The catheter was used, and then removed so the physician could treat the lesion with another method.When they went to reinsert the jetstream catheter, it would not advance over the wire at the tip end.Another of the same device was used to complete the procedure with the same wire.However, device analysis revealed foreign material in the tip of the device.
|