The bwi product analysis lab received the device for evaluation.The product investigation was subsequently completed.Device evaluation details: the product was returned to biosense webster for evaluation.Bwi then conducted visual inspection of the returned device.Visual analysis of the returned sample revealed that tip was cut.The root cause cannot be determined, it could be related to the handling of the device, however, this cannot be conclusively determined.Unit was inspected prior leaving the facility as there are functional tests and inspections at control points based on the process to avoid this type of damage from leaving the facility.A manufacturing record evaluation was performed for the finished device (b)(4) number, and no internal actions related to the reported complaint condition were identified.Although no electrical issue could be reach on the cause of the reported event; to minimize ecg noise, this guidelines should be followed: ecg noise is typically generated as the result of improper connection of the body surface ecg patch to the patient.This noise is the most significant during ablation.To resolve this situation, verify proper connection.It is recommended to turn off the notch filter for this verification.It is recommended that the observation room and the operating room both be connected to a common protective ground.Doing so prevents leakage current between various ep lab equipment and noise on ecg channels.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's reference number: (b)(4).
|