Model Number N/A |
Device Problem
Difficult to Insert (1316)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/17/2021 |
Event Type
Injury
|
Manufacturer Narrative
|
(b)(4).Concomitant medical products: 42532007101 - natural tibia cemented 5 degree stemmed left size e - 64851792.Report source: foreign country: (b)(6).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
Event Description
|
It was reported that a total knee arthroplasty was performed with persona implants and during the procedure, the persona vivacit-e ps surface didn't firmly mate with the tibia component.Subsequently, another articular surface was used to complete the procedure.There was no surgical delay.The surgical technique was utilized.Attempts have been made and no further information has been provided.
|
|
Event Description
|
No further event information available at the time of this report.
|
|
Manufacturer Narrative
|
This follow-up report is being submitted to relay additional information.Visual examination of the returned product identified the distal surface to exhibit damage (nicked, gouged) and the dove tail/locking features to be flared out.The dhr was reviewed and no discrepancies relevant to the reported event were found.Medical records were not provided.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Search Alerts/Recalls
|