ETHICON ENDO-SURGERY, LLC. HARMONIC SCALPEL HAND PIECE; INSTRUMENT, ULTRASONIC SURGICAL
|
Back to Search Results |
|
Catalog Number HP054CN |
Device Problem
Device Markings/Labelling Problem (2911)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/05/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Photo analysis: this is a analysis for an image submitted to ethicon endo surgery for evaluation.Image: the image provided by the customer is that where a manufacturing date at the package box has a typo error.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.Based on the information currently available, a product issue was identified during the investigation of the sample received.This product issue will be addressed through ethicon endo surgery quality system.Because the instrument was not returned our evaluation is limited.We value the opportunity to fully analyze the instrument upon its return.
|
|
Event Description
|
It was reported that a wrong dom was on the label.It was reported that the distributor found the dom on the english label for the package was 2021-05-94, it wasn't used on patient.
|
|
Manufacturer Narrative
|
(b)(4).Date sent: 1/24/2022.D4 batch #: v94t4x an analysis of the product could not be performed since a physical sample was not received for evaluation.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.As part of ethicon's quality process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of post market surveillance.If the product is received at a later date, the investigation will be updated as applicable.
|
|
Search Alerts/Recalls
|
|
|