ST. JUDE MEDICAL TACTICATH CONTACT FORCE ABLATION CATHETER, SENSOR ENABLED BID CURVE D-F; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FIBRILLATION
|
Back to Search Results |
|
Model Number A-TCSE-DF |
Device Problem
Contamination (1120)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 05/17/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
One bi-directional, curve d-f, tacticath sensor enabled contact force ablation catheter was received for evaluation.Microscopic inspection of the distal tip confirmed a foreign material inside the center of the distal tip.Further investigation revealed fibrous and thread like foreign material (fm).The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the noted foreign material remains unknown.
|
|
Event Description
|
This report is being submitted based on analysis which revealed foreign material in the distal tip of the catheter.
|
|
Manufacturer Narrative
|
Additional information: g3.Correction: h6 investigation conclusion code.
|
|