Catalog Number 2C8750 |
Device Problem
Device Contamination with Chemical or Other Material (2944)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/02/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Should additional relevant information become available, a supplemental report will be submitted.
|
|
Event Description
|
It was reported that a piece of plastic was observed in the y-site of clearlink system y-type blood/solution set.This issue was observed when the set was connected to a pediatric patient.There was no patient injury or medical intervention associated with this event.No additional information is available.
|
|
Manufacturer Narrative
|
The user facility submitted medwatch 0700220000-2021-8063 for this event.H10: the device was received for evaluation.Visual inspection was performed using the naked eye which observed a small clear plastic particulate in the tubing.The reported condition was verified.The cause of the condition was due to the manufacturing process.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
|