BOSTON SCIENTIFIC CORPORATION INTELLANAV MIFI OPEN-IRRIGATED; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FLUTTER
|
Back to Search Results |
|
Model Number 87047 |
Device Problems
Material Perforation (2205); Communication or Transmission Problem (2896); Temperature Problem (3022)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/12/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
It was indicated that the device will not be returned for evaluation.If there is any further relevant information obtained, a supplemental medwatch will be filed.
|
|
Event Description
|
During an redo pulmonary vein isolation procedure to treat atrial fibrillation an intellanav mifi open-irrigated was selected for use.The temperature read 68 degrees celsius in the patient and when removed from the patient the temperature dropped to 24 degrees.Further information alleges that there was an issue with shaft tears, rips, holes, or separated material.An exchange of the catheter resolved the issue.The procedure was completed without any patient complications.
|
|