ZOLL MANUFACTURING CORPORATION LIFEVEST WCD 4000 SYSTEM; WEARABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number WCD 4000
Device Problem Defibrillation/Stimulation Problem (1573)
Patient Problem Insufficient Information (4580)
Event Date 07/18/2021
Event Type  Death  
Manufacturer Narrative
Device evaluation of monitor sn (b)(4) and electrode belt sn (b)(4) has been completed.The reported problem (patient death) was confirmed.The evaluation included review of downloaded software flag files on the day of the event and incoming functional testing.During the incoming functional testing, a 1hz simulated normal sinus rhythm signal was applied to the ecg electrodes, followed by a 5hz simulated treatable arrhythmia signal which verified proper performance of the detection algorithm.During the transition to the 5hz signal, the device was confirmed to properly enter into a treatment sequence which includes a verification of the tactile vibration alarm, audio messaging, and siren alarms, as well as a test of the pulse delivery circuitry.The pulse delivery circuitry test verified proper charging of the high voltage capacitors and proper delivery of five full energy 150j biphasic pulses.The functional testing confirmed proper response button functionality, ecg acquisition, detection algorithm performance, and pulse delivery functionality.  upon further investigation of the download data, the monitor had displayed service code 104, indicating a defective sd card.A defective sd card does not affect the monitor's ability to detect and treat an arrhythmia.Unavailability of retrospective ecg recordings did not cause or contribute to the this adverse event.The sd card is a data logger that only stores the patient's retrospective continuous ecg recordings.There is no real and active patient monitoring associated with the stored retrospective ecg recordings.  the root cause of the defective sd card could not be positively identified.Device manufacture date: monitor 04/04/2014, belt 04/15/2015.
 
Event Description
A us distributor contacted zoll to report that a patient had passed away at home while wearing the lifevest on (b)(6) 2021 around 9:00 pm.It was reported that the patient was treated by the lifevest on the date of passing.A review of the downloaded patient data flag files indicate that the patient received eight unknown treatment events at 01:04:59, 01:05:30, 15:15:29, 15:16:00, 15:16:27, 15:16:56, 15:17:25, and 15:21:17 on the date of passing ((b)(6) 2021).The device was shutdown at 15:59:38.The patient reportedly passed away around 9:00 pm.The specific rhythm at the time of the treatment events is unknown.The downloaded data indicates that the device had displayed multiple service code 104 (sd card fault) flags on the days prior to the date of passing.There were no allegations of device malfunction contributing to the patient's death.The device was fully functional upon receipt.Reporting this event out of an abundance of caution as the rhythm at the time of the treatment events on the patient's date of death is unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEVEST WCD 4000 SYSTEM
Type of Device
WEARABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ZOLL MANUFACTURING CORPORATION
121 gamma drive
pittsburgh PA 15238 3495
Manufacturer (Section G)
ZOLL MANUFACTURING CORPORATION
121 gamma drive
pittsburgh PA 15238 3495
Manufacturer Contact
matthew saich
121 gamma drive
pittsburgh, PA 15238-3495
4129683333
MDR Report Key12424678
MDR Text Key270036356
Report Number3008642652-2021-07666
Device Sequence Number1
Product Code MVK
UDI-Device Identifier00855778005005
UDI-Public00855778005005
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010030
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor
Reporter Occupation Other
Type of Report Initial
Report Date 09/03/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/03/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model NumberWCD 4000
Device Catalogue Number10A0988
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/02/2021
Date Manufacturer Received08/12/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/04/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Death;
-
-