Model Number 397002-001 |
Device Problems
Computer Software Problem (1112); Mechanical Problem (1384)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/25/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
The companion 2 driver will be returned to syncardia for evaluation.The results will be provided in a follow-up mdr.(b)(4).
|
|
Event Description
|
The companion 2 driver was not supporting a patient.The customer, a syncardia certified hospital, reported that the companion 2 driver did not boot up with the proper user interface screen.
|
|
Manufacturer Narrative
|
The patient data file was reviewed and no alarms were recorded that would indicate a device malfunction.During functional testing, the key was observed to be hard to turn and had to be forced to the "on" position.With the key switch faulty and not allowing proper boot up, it is likely that the left and right embedded supervisors were not properly communicating which could lead to improper boot up issues to include a white screen as reported by the customer.The root cause of the customer-reported issue was determined to be a malfunction of the companion 2 driver key switch.Syncardia has a corrective and preventive action (capa) for the issue of companion 2 driver broken keys/key switch.Syncardia has completed its investigation and is closing this file.(b)(4) follow- up report 1.
|
|