• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER SURGICAL, INC. HANDPIECE, ELECTRIC DERMATOME

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER SURGICAL, INC. HANDPIECE, ELECTRIC DERMATOME Back to Search Results
Model Number N/A
Device Problems Degraded (1153); Electrical /Electronic Property Problem (1198); Unstable (1667)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/14/2021
Event Type  malfunction  
Manufacturer Narrative
This event has been recorded by zimmer biomet under cmp-(b)(4).Review of the most recent repair record determined that the insulation of the power cable was damaged.The plug harness was replaced and resolved the reported issue.Additionally, during inspection the motor speed was unstable, some screws were worn, the unit was out of calibration, and the position of the control bar was not correct.The motor and screws were replaced, the device was re-calibrated, and the control bar position was corrected.Device is used for treatment.A definitive root cause cannot be determined.The event is confirmed.
 
Event Description
It was reported that during post intervention, there might be a false contact of wires that seemed to come from this power cable.There were no exposed wires.Investigation found the motor speed was unstable.No adverse event was reported as a result of this malfunction.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HANDPIECE, ELECTRIC DERMATOME
Type of Device
DERMATOME
Manufacturer (Section D)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer (Section G)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer Contact
christina arnt
56 e. bell drive
warsaw, IN 46582
5745273773
MDR Report Key12434052
MDR Text Key270156676
Report Number0001526350-2021-00994
Device Sequence Number1
Product Code GFD
UDI-Device Identifier00889024375994
UDI-Public(01)00889024375994(11)190619(10)64399077
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,user facility
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 09/07/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/07/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number00882100100
Device Lot Number64399077
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/01/2021
Was the Report Sent to FDA? No
Date Manufacturer Received09/02/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/19/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
-
-