Catalog Number 03666794001 |
Device Problem
Display or Visual Feedback Problem (1184)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/19/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
The customer's meter was requested for investigation and a replacement meter was sent to the customer.Initial reporter occupation - the occupation is lay user/patient.
|
|
Event Description
|
It was reported that there was an issue with the display of coaguchek xs meter serial number (b)(4).The meter display had missing segments in the results field.There have been no misinterpreted results.The meter batteries were changed, but segments were still missing from the results field when a display check was performed.
|
|
Manufacturer Narrative
|
The customer's meter was received for investigation.The meter¿s circuit board was tested for damage or contamination and it was found the printed circuit board (pcb) was contaminated.The root cause is determined to be contamination of the contacts due to improper handling or maintenance by the customer.
|
|