Model Number 10280959
Device Problem Mechanical Problem (1384)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/31/2021
Event Type  malfunction  
Manufacturer Narrative
Siemens is conducting a thorough investigation of the reported events.As this event is under investigation, a root cause has not yet been determined.A supplement report will be filed upon completion of the investigation.
 
Event Description
It was reported to siemens that a malfunction occurred while operating the artis zeego system.During an interventional procedure, the user reported that the flat panel cannot be moved up and down.The procedure was continued and completed on an alternate system.We are unaware of any impact to the state of health of the patient involved.Siemens has requested additional information in order to conduct an investigation of the reported event.
 
Manufacturer Narrative
Siemens has completed an investigation of the reported event.The root cause was determined to be a temporary hardware error.The investigation determined a bad contact at connector x3 of the control unit of the source-to-image distance control (sid).The error was eliminated by the activities on site (e.G., pulling and plugging).Both the sid motor and the associated control unit were replaced.The system ran again as specified.The occurrence rate of the error pattern and the spare parts consumption of the components were checked.A possible error accumulation or even a systematic error, which would lead to a corrective action of the installed base, could not be determined by the investigation.After detailed investigation, the incident is not classified as a reportable event as neither serious injury, death nor an unexpected, prolonged hospitalization of the patient or any other person occurred or could be expected.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARTIS ZEEGP
Type of Device
INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM
Manufacturer (Section D)
SIEMENS HEALTHCARE GMBH- AT
siemensstrasse 1
forchheim, germany 91301
GM  91301
MDR Report Key12438921
MDR Text Key270302070
Report Number3004977335-2021-95572
Device Sequence Number1
Product Code OWB
UDI-Device Identifier04056869010083
UDI-Public04056869010083
Combination Product (y/n)N
PMA/PMN Number
K181407
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Remedial Action Repair
Type of Report Initial,Followup
Report Date 09/08/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/08/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number10280959
Was Device Available for Evaluation? Yes
Date Manufacturer Received10/13/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-