Additional information: d9, h3, h6, h10.One actual sample was received for evaluation.A visual inspection with the naked eye noted the patient adapter slipped back from the bushing/tubing.Functional testing including leak testing, clear passage testing and clamp function testing were performed with no issues noted.The reported condition was verified.The cause of the condition could not be determined.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|