Model Number L20410 |
Device Problems
Defective Device (2588); Adverse Event Without Identified Device or Use Problem (2993)
|
Patient Problem
Limb Fracture (4518)
|
Event Date 08/18/2021 |
Event Type
Injury
|
Manufacturer Narrative
|
Product complaint #(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Event Description
|
The corail broach has a square shape on the medial aspect.Comments are if the resection cut of the femur is high then this broach shape causes a fracture of the anterior medial cortex of the femur when reducing the hip after a bi polar trial construct is created.Doi: unknown, dor: unknown, unknown side.
|
|
Manufacturer Narrative
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
|
Event Description
|
There is no alleged product deficiency.
|
|
Manufacturer Narrative
|
Product complaint (b)(4).Investigation summary: no device was received for examination, therefore the reported event could not be confirmed.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot : a manufacturing record evaluation (mre), was not possible because the required lot code was not provided.
|
|