• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNCARDIA SYSTEMS, LLC SYNCARDIA FREEDOM HOSPITAL AC POWER SUPPLY; EXTERNAL POWER SUPPLY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNCARDIA SYSTEMS, LLC SYNCARDIA FREEDOM HOSPITAL AC POWER SUPPLY; EXTERNAL POWER SUPPLY Back to Search Results
Catalog Number 295400-002
Device Problem Material Integrity Problem (2978)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/06/2021
Event Type  malfunction  
Manufacturer Narrative
The freedom hospital ac power supply will be returned to syncardia for evaluation.The results will be provided in a follow-up mdr.Ce (b)(4).
 
Event Description
The freedom hospital ac power supply is a component that enables the freedom driver to be plugged into an external power source.The customer, a syncardia authorized distributor, reported that the freedom hospital ac power supply had a damaged connector.
 
Manufacturer Narrative
During visual inspection, damage to the connector was confirmed.Additionally, three of the four power supply rubber feet were missing and physical damage was observed at the strain relief connecting the power cord to the brick.During investigation testing, the power supply was found to be functional despite the observed damage.The observed damage did not affect the functionality of the power supply and was mostly likely due to rough handling.This issue will be monitored and trended as part of the customer experience process.Syncardia has completed its investigation and is closing this file.(b)(4) follow-up report 1.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCARDIA FREEDOM HOSPITAL AC POWER SUPPLY
Type of Device
EXTERNAL POWER SUPPLY
Manufacturer (Section D)
SYNCARDIA SYSTEMS, LLC
1992 e. silverlake road
tucson AZ 85713
Manufacturer (Section G)
SYNCARDIA SYSTEMS, LLC
1992 e. silverlake road
tucson AZ 85713
Manufacturer Contact
kerri hensley
1992 e. silverlake road
tucson, AZ 85713
5205451234
MDR Report Key12451262
MDR Text Key273613563
Report Number3003761017-2021-00183
Device Sequence Number1
Product Code LOZ
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
P030011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Distributor
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 03/04/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/09/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue Number295400-002
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/02/2021
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/06/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/09/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age44 YR
Patient SexFemale
-
-