H10.Additional manufacturer narrative: the device history record was reviewed and shows that this device met all manufacturing specifications for product release prior to distribution.No issues were identified that would have impacted this event.H3.Evaluation: customer report of intervalvular leak and valve sizing issues could not be confirmed through visual observations.X-ray demonstrated wireform intact.All three leaflets had serrated mechanical damage on the outflow surfaces.Suture holes were visible around the sewing ring.Sewing ring was cut around commissure 3.The external sewing ring diameter measured to be approximately 28 mm and the internal stent diameter measured to be approximately 22 mm.No intraoperative echocardiogram was provided, thus, the behavior reported in vivo could not be confirmed.A review of the manufacturing and inspection records related to this device was completed, and the results show that the device met all manufacturing specifications and inspections.The reported surgeon feels this particular valve was mis-sized by manufacturer was not confirmed.Sizing is a function of the prosthetic valve and the adjacent anatomy.Since only the valve was returned to edwards, no statement can be made concerning the sizing of the valve.Functional testing was performed in a pulse duplicator under simulated normal physiological conditions and in a steady backflow leakage tester.The leaflets all functioned well, opening and closing normally.The total regurgitant fraction (trf) of the valve as received was 8.9 percent, which is lower than the 10 percent allowed for a valve this size per iso 5840-2:2021.No visible central leakage pathway was evident during pulsatile testing.Therefore, the leakage measured in the pulsatile flow test is likely through the damaged areas of the sewing ring and/or around the valve.The reported intervalvular leak was not reproduced in the standardized pulsatile testing performed at edwards under simulated normal physiological conditions.The results from in vitro testing may be different from those obtained in the operating room due to hemodynamic and environmental differences.Per ifu, the specification for the external sewing ring diameter of 3300tfx23mm is 28mm, and the internal stent diameter is 22mm.Measurements performed in product evaluation align with nominal measurements provided in the ifu.Therefore, the customer allegation that the valve was not manufactured to the correct size was not confirmed.A definitive root cause for the reported intervalvular leak in vivo could not be conclusively determined, however it may have been due to patient and/or procedural factors.Based on the information available, a manufacturing defect has not been confirmed.
|