• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOTRONIK SE & CO. KG EDORA 8 DR-T; PACEMAKER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOTRONIK SE & CO. KG EDORA 8 DR-T; PACEMAKER Back to Search Results
Model Number 407145
Device Problem Ambient Noise Problem (2877)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/06/2021
Event Type  malfunction  
Event Description
Patient presented for an rv lead revision.When new rv lead was connected to can, both channels had extreme noise on them.Ra lead was fine prior to procedure.Physician disconnected both leads, cleaned off the end terminal and reconnected.There was still noise on both channels.This device was explanted and a new one was implanted.There was no noise on the new can.
 
Manufacturer Narrative
We received your event description for the above mentioned device and would like to thank you for supporting our post-market surveillance.As of today, the medical device is not available for analysis, therefore the device itself could not be investigated.The information you provided has been entered into our quality system as a complaint.These types of complaints are used to evaluate systems and device performance throughout our organization and help to maintain and improve the performance of our devices.Should additional relevant information or the device itself become available, the investigation will be updated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EDORA 8 DR-T
Type of Device
PACEMAKER
Manufacturer (Section D)
BIOTRONIK SE & CO. KG
woermannkehre 1
berlin 12359
MDR Report Key12474413
MDR Text Key271476321
Report Number1028232-2021-05125
Device Sequence Number1
Product Code NVZ
UDI-Device Identifier04035479147648
UDI-Public04035479147648
Combination Product (y/n)N
PMA/PMN Number
P950037
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 09/06/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/15/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/31/2022
Device Model Number407145
Device Catalogue NumberSEE MODEL NO.
Was Device Available for Evaluation? No
Date Manufacturer Received09/16/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Hospitalization;
Patient Age69 YR
-
-