SMITH & NEPHEW, INC. GENII 7.5MM RESUR PAT TRL 29MM; PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
|
Back to Search Results |
|
Model Number 71928373 |
Device Problems
Crack (1135); Material Discolored (1170); Scratched Material (3020)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 06/17/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
The associated device, intended for use in treatment, was returned and evaluated.A visual inspection of the returned genii 7.5mm resur pat trl 29mm confirms the device is cracked on one side with discoloration in the plastic.The device also has extensive wear usage.A review of the manufacturing records did not reveal any manufacturing or material abnormalities that could have caused or contributed to the reported incident.A review of complaint history did not reveal additional complaints for the listed batch.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.
|
|
Event Description
|
It was reported that a genii 7.5mm resur pat trl 29mml has a cut in it from saw blade.As this was noticed upon field inspection, there was no patient involvement.The investigation found that it was cracked.
|
|
Search Alerts/Recalls
|
|
|