The patient was undergoing a coil embolization procedure in the superior mesenteric artery using a lantern delivery microcatheter (lantern), a guide catheter and a guidewire.During the procedure, the physician successfully implanted ten coils in the target vessel using the lantern.Subsequently, the physician removed the lantern from the patient to reposition the guide catheter.Upon reinserting the lantern into the patient, the physician noticed a pin hole leak on the distal end of the lantern.Therefore, the lantern was removed.It was reported that the physician was slightly rough with the lantern.The procedure was completed using another lantern and the same guide catheter.There was no report of an adverse effect to the patient.
|
Evaluation of the returned lantern revealed a fracture on the proximal shaft.If the device is manipulated against resistance or is mishandled at extreme angles during use, damage such as a kink and subsequent fracture may occur.This damage may be the reported pin hole leak as there was no hole observed on the distal end of the catheter.Further evaluation revealed a kink on the proximal shaft and an ovalization on the distal shaft.This damage may be incidental to the reported complaint and may have occurred during packaging for return to penumbra.Penumbra catheters are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.H3 other text : placeholder.
|