• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TOSOH HI-TEC, INC. AIA-360; FLUOROMETER, FOR CLINICAL USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TOSOH HI-TEC, INC. AIA-360; FLUOROMETER, FOR CLINICAL USE Back to Search Results
Model Number AIA-360 IMMUNOASSAY ANALYZER
Device Problems Break (1069); Material Twisted/Bent (2981)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/19/2021
Event Type  malfunction  
Manufacturer Narrative
A field service engineering (fse) was at the customer's site to address reported event.The customer reported getting errors 2012 and 4004.Fse confirmed the complaint by reviewing the error logs and reproduced errors by attempting to move the turn table manually and the carousel would not move.Fse replaced the bent sample needle.Fse also found the sample detection spring was broken and not touching the case of the eki board.The broken piece was inside of the carousel pitch sensor and jamming it, preventing free movement of the carousel.Fse removed the broken piece of the spring which freed the carousel.Fse cleaned the carousel, z axis of the sample arm and verified all alignments of the sample nozzle.Fse validated the analyzer by successfully performing quality control run without error and within acceptable range.No further action required by field service.The aia-360 analyzer is functioning as expected.A 13-month complaint history review and service history review through aware date of event for similar complaints was performed for serial number (b)(4).There were two (2) similar complaints identified during the searched period, which includes this event.The aia-360 operator's manual under chapter 7, section 7.1: list of error messages states the following: air detected [diluent] is generated when there is no contact with the liquid after diluent suction.The operator is instructed to contact the service department.Turn table slip is generated when the turntable motor slipped.The operator is instructed to turn the power off and on again.If this problem reoccurs, contact the service department.The most probable cause of the reported events were due bent sample nozzle and a broken sample detection spring.
 
Event Description
A customer reported getting error message 2012 ¿air detected [diluent]¿ on the aia-360 analyzer.While troubleshooting with the customer over the phone, the technical support specialist (tss) noted the sample nozzle was bent.Analyzer is down.A field service engineer (fse) was dispatched to address the reported event, which resulted in a delayed reporting of patient samples for beta human chorionic gonadotropin (bhcg) and cardiac troponin i (ctnl2).There is no indication of any patient intervention or adverse health consequences due to the delay in reporting of patient results.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AIA-360
Type of Device
FLUOROMETER, FOR CLINICAL USE
Manufacturer (Section D)
TOSOH HI-TEC, INC.
1-37, fukugawa minami-machi
shunan, yamaguchi 74600 42
JA  7460042
Manufacturer (Section G)
TOSOH HI-TEC, INC.
1-37, fukugawa minami-machi
shunan, yamaguchi 74600 42
JA   7460042
Manufacturer Contact
bernadette o connell
shiba-koen first building
3-8-2 shiba
minato-ku 10586-23
JA   1058623
6506368143
MDR Report Key12485687
MDR Text Key271900022
Report Number3004529019-2021-00031
Device Sequence Number1
Product Code KHO
UDI-Device Identifier04560189282896
UDI-Public04560189282896
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K971103
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other Health Care Professional
Remedial Action Repair
Type of Report Initial
Report Date 09/17/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/16/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberAIA-360 IMMUNOASSAY ANALYZER
Device Catalogue Number019945
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/19/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/01/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-