HOYA CORPORATION PENTAX TOKYO OFFICE PENTAX; VIDEO PROCESSOR Back to Search Results
Model Number EPK-I7000
Device Problem Loose or Intermittent Connection (1371)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/20/2021
Event Type  malfunction  
Manufacturer Narrative
If additional information becomes available, a supplemental report will be filed with the new information.This device is not marketed in us, therefore 510k is not applicable.
 
Event Description
During connecting the scope and processor, the scope lock lever loosed and could not lock the scope.The time of event is unknown.There was no report of patient harm.
 
Manufacturer Narrative
Evaluation summary: the screws loosed which caused scope lock lever loosed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENTAX
Type of Device
VIDEO PROCESSOR
Manufacturer (Section D)
HOYA CORPORATION PENTAX TOKYO OFFICE
tsutsujigaoka 1-1-110
akishima-shi, tokyo 196-0 012
JA  196-0012
Manufacturer (Section G)
HOYA CORPORATION PENTAX TOKYO OFFICE
tsutsujigaoka 1-1-110
akishima-shi, tokyo 196-0 012
JA   196-0012
Manufacturer Contact
william goeller
3 paragon drive
montvale, NJ 07645
8004315880
MDR Report Key12485812
MDR Text Key280803695
Report Number9610877-2021-00987
Device Sequence Number1
Product Code PEA
Combination Product (y/n)N
Reporter Country CodeCH
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 12/21/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/16/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberEPK-I7000
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/24/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-