Model Number 176630 |
Device Problem
Component Missing (2306)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 05/09/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Evaluation summary: medtronic conducted an investigation based upon all information received.The device and a photo were available for evaluation.Visual inspection noted the trip lever was out of position.It was reported that the instrument had less than the expected amount of clips.The reported issue was confirmed.A review of the device history record for the clip applier indicates the product was released meeting all medtronic quality release specifications at the time of manufacture.However, a manufacturing fault was identified during product analysis.The root cause of the observed condition was determined to be a result of a manufacturing activity.The trip lever out of position preventing the clips from loading properly.A process improvement has been initiated to prevent this condition from recurring.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
According to the reporter, pre operatively, there was no clip inside from the beginning.There was no patient involvement.
|
|