The catalog number identified has not been cleared in the us, but is similar to the lifestent 5f vascular stent system products that are cleared in the us.The pro code and 510k number for the lifestent 5f vascular stent system products is identified.Manufacturing review: a review of manufacturing records was not performed, as additional complaints have not been reported for this lot.Based on the information available it is not reasonably suggested that a manufacturing process may have caused or contributed to the reported issue.However, the lot history records of this lot were reviewed with special attention to the manufacturing and inspection of this product and the product was found to have met the specification prior to shipment.Investigation summary: the sample was returned for evaluation.The investigation of the returned delivery system confirmed that the user could not deploy the stent.A force transmitting adhesive joint was found loose which made a deployment impossible.An indication for a manufacturing related issue could not be found.Based on the information available the investigation is closed with confirmed result.A definite root cause for the reported event could not be determined.Labeling review: in reviewing the relevant labeling for this product, the potential issue was found addressed.The instructions for use state: 'if excessive force is felt during stent deployment, do not force the delivery system.Remove the delivery system and replace with a new unit.' under materials required the instructions for use state: '5f (1.67 mm) or larger introducer sheath (¿) 0.014 inch (0.36 mm) - 0.035 inch (0.89 mm) diameter guidewire.' regarding pre dilation the instructions for use state: 'pre-dilatation of the lesion with a balloon dilatation catheter is recommended.' (expiry date: 02/2021).
|