Model Number GF-UCT180 |
Device Problems
Break (1069); No Display/Image (1183); Device Contamination with Chemical or Other Material (2944)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The subject device was not returned to omsc for evaluation but was returned to olympus (b)(4).(b)(4) checked the subject device and found the reported phenomena.The exact cause of the reported event could not be conclusively determined at this time.If additional information is received, this report will be supplemented.
|
|
Event Description
|
Olympus medical systems corp.(omsc) was informed that during the incoming inspection for repair at the service department of olympus (b)(4), it was found that the distal end cover had crack and there was foreign material on the forceps elevator.The occurrence date of the event is unknown.There was no report of patient injury associated with the event.
|
|
Manufacturer Narrative
|
This supplemental report is being submitted to provide additional information.Omsc reviewed the manufacturing history (dhr) of the subject device and confirmed no irregularity.The exact cause of the reported event could not be conclusively determined.Omsc considered that the reported phenomena (the foreign material on the forceps elevator, the crack of the distal end) might be subsequent events.If additional information becomes available, this report will be supplemented.
|
|
Manufacturer Narrative
|
This report is being supplemented to provide information that was inadvertently left out of the initial medwatch and the supplemental medwatch.This mdr is being submitted as part of a retrospective review and remediation effort based on enhancements made to the company¿s mdr and complaint handling processes.Capas have been opened to manage the actions that are being taken to remediate this issue and ensure any required mdr reporting is completed.This report is being submitted for the reportable malfunction of no ultrasound image reported by the customer.The subject device was returned to an olympus service center for evaluation.Upon inspection and testing of the returned device, the customer's complaint was not confirmed.A review of the device history record found no deviations that could have caused or contributed to the reported issue.The device met all specifications at the time of shipment.Based on the results of the legal manufacturer's investigation, although a conclusive root cause could not be determined, it is likely an ultrasound image was not displayed due to following: ultrasound image was lost due to ultrasound probe breakage.Ultrasound image was lost due to the ultrasound cable disconnection.Ultrasound image was lost due to the failure of the pc board.The following information is stated in the ifu (instructions for use) which may help to prevent the issue: ¿ inspection of the endoscopic system, warnings and cautions or precautions, and storage of the ultrasonic cable.¿ olympus will continue to monitor field performance for this device.
|
|
Manufacturer Narrative
|
This report is being submitted to correct the legal manufacturer¿s contact information and facility registration number.The facility registration number is 3002808148.
|
|